MLC Transmission and Dosimetric Leaf Gap Measurement Using CU Values from Integrated Images of Varian Vital Beam LINAC

Authors

  • Md. Motiur Rahman

  • Md Fajle Rabby

  • Mahmuda Akter

  • Rubel Ahmed

  • MMH Bhuiyan

  • MKA Khan,

Keywords:

electronic portal imaging device (EPID), VMAT, dosimetric leaf gap (DLG)

Abstract

Modern dynamic radiotherapy techniques require more precision and needroutine checking. However, presently available methods for Dosimetric Leaf Gap (DLG) and MLC transmission factor are highly timeconsuming, so theyare not feasible for busy radiotherapy centers. Thisanalytical and observational study tried to determine the most straightforward way of simultaneously determining the transmission and DLG in the shortest time. We measured all data at 5cm depth using a Varian Vital Beam LINAC, newly installed at TMSS Cancer Center, Bogura, Bangladesh. Weused three significant pieces of equipment in our experiment: an integrated Electronics Portal Imaging Device (EPID), a solid water phantom and a farmer-type ion chamber (Model: IBA FC65-P). The Eclipse™ TPS was also used to make the QA plan and to analyze the images. As per Varian machine recommendation, the transmission factor and DLG will be less than 2% and 2mm, respectively. In our measurement, using a Farmer-type ion chamber, we got the value of transmission factor of 1.46% and 1.69% and DLG of 0.98 mm and 1.13 mm for 6MV and 10MV photons, respectively. Using EPID a-Si1200, we got the transmission of 1.23% and 1.63% and DLG of 0.85 mm and 1.08 mm for 6MV and 10MV photons, respectively. MLC transmission is in good agreement with ion chamber-based dosimetry with established data. It is a little lower for the 6MV photon EPID-based result, that is, maybe for the inhomogeneity characteristics of the photon fluence in low energies and negligible. For 10MV or higher energy photons, it is almost similar.

How to Cite

MLC Transmission and Dosimetric Leaf Gap Measurement Using CU Values from Integrated Images of Varian Vital Beam LINAC. (2023). Global Journal of Medical Research, 23(17), 29-36. https://medicalresearchjournal.org/index.php/GJMR/article/view/102343

References

J. Adamson, Q. Wu (2012) Adamson J, Wu Q. Independent verification of gantry angle for pre-treatment VMAT QA using EPID. 57(20), 6587. https://doi.org/10.1088/0031-9155/57/20/6587

Fuangrod T, Woodruff HC, Van E, McCurdy BM, Kuncic Z, O'Connor DJ, Greer PB (2013) A system for EPID‐based real‐time treatment delivery verification during dynamic IMRT treatment. 40(9). https://doi.org/10.1118/1.4817484

Otto K (2008) Volumetric modulated arc therapy: IMRT in a single gantry arc Medical physics. 35(1), 310-317. https://doi.org/10.1118/1.2818738

Nicolini G, Clivio A, Cozzi L, Fogliata A, Vanetti E (2011) On the impact of dose rate variation upon RapidArc® implementation of volumetric modulated arc therapy Medical physics. 38(1), 264-271. https://doi.org/10.1118/1.3528214

Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, Serago C, Aguirre F, Ma L, Arjomandy B, Liu C (2009) Task Group 142 report: Quality assurance of medical accelerators a Medical physics. 36(9Part1), 4197-4212. https://doi.org/10.1118/1.3190392

Koren Smith, Peter Balter, John Duhon, Gerald White, David Vassy, Robin Miller, Christopher Serago, Lynne Fairobent (2017) AAPM Medical Physics Practice Guideline 8 a.: linear accelerator performance tests Journal of applied clinical medical physics. 18(4), 23-39. https://doi.org/10.1002/acm2.12080

Ezzell GA, Galvin JM, Low D, Palta JR, Rosen I, Sharpe MB, Xia P, Xiao Y, Xing L, Yu CX (2003) Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee Medical physics. 30(8), 2089-2115. https://doi.org/10.1118/1.1591194

Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, Molineu A, Palta JR, Ramsey CR, Salter BJ, Shi J (2009) IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 Medical physics. 36(11), 5359-5373. https://doi.org/10.1118/1.3238104

Shende R, Patel G (2017) Validation of Dosimetric Leaf Gap (DLG) prior to its implementation in Treatment Planning System (TPS): True Beam™ millennium 120 leaf MLC Reports of Practical Oncology and Radiotherapy. 22(6), 485-494. https://doi.org/10.1016/j.rpor.2017.09.001

B. McCurdy, K. Luchka, S. Pistorius (2001) Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device Medical physics. 28(6), 911-924. https://doi.org/10.1118/1.1374244

Varian Associates Inc Palo Alto Ca (2011) Image Acquisition System- 3 Reference Guide.

Vieira SC, Dirkx ML, Pasma KL, Heijmen BJ (2002) Fast and accurate leaf verification for dynamic multileaf collimation using an electronic portal imaging device Medical physics. 29(9). https://doi.org/10.1118/1.1501141

K. Pasma, M. Kroonwijk, J. Boer, A. Visser, B. Heijmen (1998) Accurate portal dose measurement with a fluoroscopic electronic portal imaging device (EPID) for open and wedged beams and dynamic multileaf collimation. 43(8). https://doi.org/10.1088/0031-9155/43/8/004

J. Chavaudra, A. Bridier (2001) Definition of volumes in external radiotherapy: ICRU reports 50 and 62 Cancer radiotherapie: journal de la Societe francaise de radiotherapie oncologique. 5(5), 472-478. https://doi.org/10.1016/s1278-3218(01)00117-2

Van J (2005) Advances in modern radiation therapy. https://medicalphysics.org/documents/VanDyk1_Ch8.pdf

M. Isono, Y. Akino, H. Mizuno (2020) Inter-unit variability of multi-leaf collimator parameters for IMRT and VMAT treatment planning: a multiinstitutional survey Journal of radiation research. 61(2), 307-313.

Rahman MM, Shamsuzzaman M, Bhuiyan MM (2022) Development of spreadsheet for rapid assessment of therapeutic radiation dose delivery with electron and photon beams at various energies. 13(1), 8-12. https://doi.org/10.15406/jcpcr.2022.13.00479

M. Rahman, M. Shamsuzzaman (2021) Dosimetric characterization of medical linear accelerator Photon and Electron beams for the treatment accuracy of cancer patients World Journal of Advanced Engineering Technology and Sciences. 3(01), 41-59. https://doi.org/10.30574/wjaets.2021.3.1.0046

MLC Transmission and Dosimetric Leaf Gap Measurement Using CU Values from Integrated Images of Varian VitalBeam LINAC.

Published

2023-05-29

How to Cite

MLC Transmission and Dosimetric Leaf Gap Measurement Using CU Values from Integrated Images of Varian Vital Beam LINAC. (2023). Global Journal of Medical Research, 23(17), 29-36. https://medicalresearchjournal.org/index.php/GJMR/article/view/102343